bioZen 2.6 µm WidePore C4, LC Column 100 x 4.6 mm, Ea

Бренд: Phenomenex

Minimize your priming with our BioTi Biocompatible hardware

bioZen Titanium BioTi Hardware

Columns:
bioZen 1.8 μm SEC-3
Dimension:
150 x 4.6 mm
Mobile Phase:
50 mM Dipotassium Phosphate + 100 mM Sodium Sulfate, pH 5.0
Flow Rate:
0.3 mL/min
Temperature:
Ambient
Detection:
UV @ 280 nm
Sample:
1. γ-Globulin, 5 mg/mL 2. Ovalbumin, 1 mg/mL

Diverse mAb Comparison with Chromatographic Performance Suitable for Intact MS Analysis

  • High throughput and single method for multiple mAb analysis
  • High temperature leads improved peak shape for all mAbs
  • Formic acid mobile phase ensures sensitivity for intact mass by high resolution MS
LC Conditions
Columns:
bioZen 2.6 µm WidePore C4
Dimension:
100 x 2.1 mm
Part No.:
00D-4786-AN
Mobile Phase:
A: 0.1 % Formic acic in Water B: 0.1 % Formic acic in Acetonitrile
Gradient:
10-90 % B in 4 minutes
Flow Rate:
0.3 mL/min
Temperature:
80 °C
Detection:
UV @ 280 nm
Sample:
mAbs, Various (1 mg/mL)

Improved peak shape for intact mass applications by high resolution MS.

Sample Retention Time (min) Width @ 50%
Rituximab 3.580 0.0233
Infliximab Biosimilar 3.606 0.0272
Cetuximab 3.696 0.0270
Infliximab 3.615 0.0222

Temperature Effects on Intact and Subunit Analysis

  • Displays improved temperatures for Intact Influximab and Reduced NIST mAb at 70, 80, and 90 ºC
  • 80 ºC exhibits narrowest and most ideal peak shape for both the intact and subunit analysis
LC Conditions
Columns:
bioZen 2.6 µm WidePore C4
Dimension:
100 x 2.1 mm
Part No.:
00D-4786-AN
Mobile Phase:
A: 0.1 % Formic acic in Water B: 0.1 % Formic acic in Acetonitrile
Gradient:
10-90 % B in 4 minutes
Flow Rate:
0.3 mL/min
Temperature:
80 °C
Detection:
UV @ 280 nm
Sample:
mAbs, Various (1 mg/mL)
Артикул: 00D-4786-E0 Категория: Бренд: