Analytical LC
Clarity Oligo-RP™
Clarity Oligo-RP columns have been specifically designed for the reversed phase purification of oligonucleotides with balanced hydrophobicity and polar selectivity. Available in 3, 5, and 10 µm particles, Clarity Oligo-RP allows for high capacity analysis and purification of oligos.
A Highly Selective Analysis
|
Column:
Clarity 3 µm Oligo-RP C18
|
|
Dimension:
50 x 4.6 mm
|
|
Part No.:
00B-4441-E0
|
|
Mobile Phase:
A: 50 mM TEAA pH 7.5 B: Methanol
|
|
Gradient:
10 % to 45 % B in 30 minutes
|
|
Flow Rate:
1 mL/min
|
|
Detection:
UV @ 260 nm
|
|
Sample:
1. 40nt DNA with sequence CTTCTGAACAGTT-GATCTATG-CACTTCAGACTTATGATCA (2.5 μg)
2. 39nt DNA with sequence TTCTGAACAGTTGATCTAT-GCACTTCAGACTTATGATCA (2.5 μg)
|
Clarity Oligo-RP successfully separates a 40mer from a 39mer DNA oligonucleotide due to its excellent efficiency and resolving power.
Purification Clarity™
Easily Scale from HPLC to Preparative Purification.
|
Column:
Clarity 3 µm Oligo-RP C18
|
|
Dimension:
(A) 50 x 10.0 mm (B) 50 x 4.6 mm
|
|
Mobile Phase:
A: 50 mM TEAA pH 7.5/ 5 % Acetonitrile B: Methanol
|
|
Gradient:
10 % to 60 % B in 20 minutes
|
|
Flow Rate:
(A) 4.7 mL/ min
(B) 1.0 mL/ min
|
|
Detection:
UV @ 280 nm
|
|
Sample:
20nt DNA
|