This system achieves excellent sensitivity with an SN ratio of 60,000:1, high resolution at 0.25 cm-1, and high-speed scanning capable of 20 spectra/second. The performance of medium and higher end models is supported by high reliability including advanced dynamic alignment and an interferometer with a dehumidifier. This is compatible with applications active in a variety of circumstances, with a library of approximately 12,000 spectra and data analysis programs for contaminant analysis, and time course and rapid scan programs for reaction tracking.
IRTracer-100
| Interferometer |
Michelson interferometer (30° incident angle) Equipped with Advanced Dynamic Alignment system Sealed interferometer with Automatic Dehumidifier |
|---|---|
| Beam splitter |
Germanium-coated KBr for Middle IR (Standard) Germanium-coated CsI for Middle/Far IR (Optional) Silicon-coated CaF2 for Near IR (Optional) |
| Light source |
High-energy ceramic for Middle/Far IR (Standard) Tungsten lamp for Near IR (Optional) |
| Detector |
DLATGS detector with temperature control for Middle/Far IR (Standard) T2SL detector with liquid nitrogen cooling for Middle/Near IR (Optional) InGaAs detector for Near IR (Optional) |
| Wavenumber range |
7,800 to 350 cm-1 (Standard) 12,500 to 240 cm-1 (Optional) |
| Resolution |
0.25, 0.5, 1, 2, 4, 8, 16 cm-1 (Middle/Far IR) 2, 4, 8, 16 cm-1 (Near IR) |
| Dimensions |
600 (W) x 665 (D) x 295 (H) mm |
| Weight |
47 kg |
| S/N ratio |
60,000:1 or higher |
| Mirror speed |
Standard: 2, 2.8, 5, 9 mm/sec Optional: 10, 20, 30, 40 mm/sec (Rapid scan) |
Downloads
Download the latest brochure.