Selectivity for PAH Compounds
The Kinetex PAH is specifically developed to be the first Core-Shell based LC product for PAH analysis. Perform high resolution PAH separations with fast run times on both HPLC or UHPLC instrumentation with the Kinetex 3.5 µm PAH column.
QC Tested for PAHs (EPA 610)
Ultimately, Kinetex PAH will provide expanded resolution with chemical selectivity specifically for PAHs. It will also increase HPLC/UHPLC throughput and sensitivity with core-shell technology.
EPA 610 – PAH Analysis
| Column: Kinetex 3.5 µm PAH |
| Dimension: 100 x 4.6 mm |
| Part No.: 00D-4764-E0 |
| Mobile Phase: A: Water B: Acetonitrile |
| Gradient: Time (min) 0 7 8 9 12 |
% B 50 100 100 50 50 |
| Flow Rate: 1.2 mL/min |
| Backpressure: 136 Bar |
| Temperature: 35 ˚C |
| Detection: UV @ 292 nm |
| Samples: 1. Naphthalene 2. Acenaphthylene 3. Acenaphthene 4. Fluorene 5. Phenanthrene 6. Anthracene 7. Fluoranthene 8. Pyrene 9. Benzo[a]anthracene 10. Chrysene 11. Benzo[b]fluoranthene 12. Benzo[k]fluoranthene 13. Benzo[a]pyrene 14. Dibenz[a,h]anthracene 15. Benzo[g,h,i]perylene 16. Indeno[1,2,3-cd]pyrene |