Kinetex 5 µm F5 100 Å, LC Column 50 x 2.1 mm, Ea

Бренд: Phenomenex

«`html

F5 — A Novel Selectivity For Tough Separations

This pentafluorophenyl propyl column provides a very high degree of steric selectivity to separate structural isomers. The electronegative fluorine groups offer high selectivity for cationic compounds.

Dependable F5 Selectivity

While older pentafluorophenyl phases (PFP, PFPP, F5, etc.) are based on existing bonding techniques and technologies that promote irreproducibility. The Kinetex F5 was meticulously designed by Phenomenex R&D and its customers, to provide consistently accurate and high performance results.

Columns: Kinetex 2.6 µm F5 Ascentis Express 2.7 µm F5
Dimension: 50 x 4.6 mm
Mobile Phase: A: Water with 0.1% Formic Acid B: Acetonitrile with 0.1% Formic Acid
Gradient: 5-95 % B over 5 minutes
Flow Rate: 1.85 mL/min
Temperature: Ambient
Detection: UV @ 254 nm
Sample: 1. Uracil 2. Pindolol 3. Chlorpheniramine 4. Nortriptyline 5. 3-Methyl-4-Nitrobenzoic acid 6. 5-Methyl Salicyl Aldehyde 7. Hexaphenone

Take Advantage of Multiple Interactive Mechanisms

The multiple interactive mechanisms of the Kinetex F5 (pentafluorophenyl ) column successfully separate methoxybenzene isomers, while the Kinetex C18 column, which has minimal bonding interactions, cannot separate the methoxybenzene isomers. This demonstrates that columns that rely primarily on hydrophobic interactions may not be the first choice for the separation of isomeric compounds and a column with multiple interactive mechanisms may be required.

Conditions for all columns:
Columns: Kinetex 2.6 µm F5 Kinetex 2.6 µm C18
Dimension: 150 x 4.6 mm
Mobile Phase: A: Water with 0.1 % TFA (Trifluoroacetic acid) B: Acetonitrile
Isocratic: A/B (65:35)
Flow Rate: 1 mL/min
Temperature: Ambient
Detection: UV @ 254 nm
Sample: 1. 1,2,3-Trimethoxybenzene 2. 1,2-Dimethoxybenzene 3. 1,2,4-Trimethoxybenzene 4. 1,4-Dimethoxybenzene 5. Methoxybenzene 6. 1,3-Dimethoxybenzene

«`

Артикул: 00B-4724-AN Категория: Бренд: