The LC-2030C NT is an integrated high-performance liquid chromatography system which prioritizes easy-to-use features for a smooth analysis experience. The intuitive touch-screen interface and stable column performance over many continuous injections ensure reliable data every time, whether you are an HPLC beginner or an expert.
Column replacement is a breeze with the Shim-pack™ NT-ODS slide-in column. Simply slot the monolithic column into place and it connects automatically, with no need for any tools and no risk of human error.
Remote data monitoring allows more flexibility in scheduling and work from home possibilities.
Pump
Degassing unit |
Five Lines: Mobile phase 4 + Rinse solution 1 (Volume 400 µL) |
---|---|
Pumping method |
Parallel-type double plunger |
Pulsation |
≤ 0.1 MPa (1.0 mL/min,10 MPa, Water) |
Flow rate setting range |
0.0001 to 10 mL/min |
Flow rate accuracy |
≤ ±1% or ±2 µL/min, below whichever is greater (0.01 to 2 mL/min, Specified conditions) |
Flow rate precision |
≤ 0.06 %RSD or ≤ 0.02 minSD, below whichever is greater |
Configuration |
Four-solvent low-pressure gradient |
Gradient / range of set concentrations |
0 to 100%, in 0.1% steps |
Gradient / concentration accuracy |
±0.5% (0.1 to 2 mL/min, 1 to 20 MPa, Specified conditions) |
Gradient / concentration precision |
±0.1% (1 mL/min, 10 MPa, Specified conditions) |
Maximum pressure |
Packed column side : 34 MPa (0.0001 to 5 mL/min) , 22 MPa (5.0001 to 10 mL/min) |
System Delay Volume |
approximately 615 µL (Packed column side) |
Autosampler
Injection method |
Total-volume sample injection |
---|---|
Injection volume accuracy |
±1% (50 µL, N=10) |
Injection volume setting range |
0.1 to 100 µL (Option:0.1 to 50 µL) |
Injection volume reproducibility |
RSD ≤0.20% (5.0 to 100 µL), RSD ≤0.25% (2.0 to 4.9 µL) |
Cross-contamination |
0.0025% (Caffeine, Specified conditions) |
Injection cycle time |
Min. 14 sec (Specified conditions) |
Samples for processing |
336 (1 mL), 216 (1.5 mL), 112 (4 mL), 4 (MTP/DWP) |
Sample cooler |
4 to 45°C |
Injection linearity |
≥0.9999 (1 to 100 µL, Specified conditions) |
Column Oven
Heating and cooling method |
Forced air circulation method |
---|---|
Containable column size |
Slide-in column |
Temperature control range |
Room temperature −5 to 60°C, |
Temperature control precision |
±0.1°C |
Temperature stability |
±0.8°C (Specified conditions) |
Flow rate switching valve |
Not available |
PDA Detector
Wavelength range |
190 to 800 nm |
---|---|
Spectral resolution |
1.4 nm (Specified condition) |
Slit width |
1.2 nm, 8 nm |
Device resolution |
0.6 nm/pixel |
Number of photodiode array elements |
1024 |
Wavelength accuracy |
≤ ±1 nm |
Noise level |
≤ ±3 × 10−6 AU (250 nm, reference: 350 nm, Specified conditions) |
Drift |
≤ 500 × 10−6 AU/h (250 nm, reference: 350 nm, Specified conditions) |
Linearity |
Up to 2 AU (5%) |
Sampling rate |
Up to 100 Hz |
Light source |
Deuterium (D2) lamp (Standard), tungsten (W) lamp (option) |
Flow Cell |
8 µL (10mm, TC), 12 MPa |
Option cell |
Conventional: 10 µL (10 mm, TC), |
Miscellaneous
Dimensions |
W410×H605×D500 mm |
---|---|
Weight |
67 kg |
Available pH range |
1 to 13 |
Materials for parts in contact with liquids |
Stainless steel (SUS316L, SUS316), FEP, PEEK, PTFE, perfluoroelastomer, ruby, |
Workstation |
LabSolutions™ LC/GC Ver. 5.92 or later, |
Downloads
Download the latest brochure.
Ion Chromatography Solutions for Environmental Analysis [ PDF / 6.8MB ]
LC-2030C NT [ PDF / 1.03MB ]