Long Column Lifetimes and Excellent Performance
Ultra-pure, metal-free silica (99.99 (= 99.99 %) purity) is the backbone of all Luna material. The resulting high quality particles have a surface smoothness, pore structure, and pore consistency to ensure a more uniform particle shape and greater reproducibility.
Superior Particle Smoothness
Agilent Technologies® ZORBAX®
Waters® Symmetry®
Phenomenex Luna
Increase MS Sensitivity and Retention for Polar Compounds
Hydrophilic Interaction Liquid Chromatography (HILIC) is a separation mode where the partitioning of polar solutes from the high concentration‚ water-miscible‚ organic mobile phase into the hydrophilic surface environment creates separations. Polar solutes exhibit increased retention and elute in the order of increasing hydrophilicity.
Water Soluble Vitamins
|
Column:
Luna 5 μm HILIC
|
|
Dimension:
150 x 4.6 mm
|
|
Mobile Phase:
Acetonitrile/100 mM Ammonium Formate, pH 3.2 (90:10)
|
|
Flow Rate:
1.0 mL/min
|
|
Temperature:
Ambient
|
|
Detection:
UV @ 260 nm
|
|
Sample:
1. PABA 2. Nicotinamide 3. Riboflavin 4. Nicotinic Acid 5. Pyridocine 6. Thiamine
|