Greater Retention and Better Results
Higher efficiency levels in combination with excellent stationary phase coverage and greater particle inert ness, translates to improved separation power for you. Now you can utilize the greater retention of Luna Omega C18 to tackle both easy and difficult separations.
Consistent and Reliable C18 Results
Luna Omega C18
|
Columns:
Luna Omega 1.6 µm C18
|
|
|
Dimension:
50 x 2.1 mm
|
|
|
Part No.:
00B-4742-AN
|
|
|
Mobile Phase:
A: Water B: Acetonitrile
|
|
|
Gradient:
Time (min)
0 6 6.01 8
|
% B
20 60 20 20
|
|
Flow Rate:
0.4 mL/min
|
|
|
Temperature:
Ambient
|
|
|
Detection:
UV @ 220 nm
|
|
|
Samples:
1. Estriol 2. Prednisolone 3. Hydrocortisone 4. Cortisone 5. Cortisone Acetate 6. 21-Hydroxycortisone 7. 17-Hydroxycortisone 8. Deoxycorticosterone
|
|
All Purpose HPLC and UHPLC Selectivity
Phenols
|
Columns:
Luna Omega 1.6 µm C18
|
|
|
Dimension:
50 x 2.1 mm
|
|
|
Part No.:
00B-4742-AN
|
|
|
Mobile Phase:
A: 0.1 % Formic Acid in Water B: 0.1 % Formic Acid in Acetonitrile
|
|
|
Gradient:
Time (min)
0 6 7 7.1 9
|
% B
5 50 50 5 5
|
|
Flow Rate:
0.4 mL/min
|
|
|
Temperature:
22 °C
|
|
|
Detection:
UV @ 270 nm
|
|
|
Samples:
1. 3-Hydroxyphenol 2. Phenol 3. 4-Nitrophenol 4. 4-Methylphenol 5. 2-Methylphenol 6. 2,4-Dimethylphenol 7. 1-Naphthol
|
|