Upgrade to Zebron from traditional phases used for PAHs:
Say Hello to the Zebron ZB-PAH-CT GC Column
Designed to move conventional PAH testing to the exceptional, Zebron GC columns come to life through a coupling of innovative spirit and technical excellence. The Zebron ZB-PAH-EU and ZB-PAH-CT offer unparalleled performance through the power of targeted selectivity when analyzing Polycyclic Aromatic Hydrocarbons (PAHs).
Excellent Resolution of Chrysene and Triphenylene
We designed the Zebron ZB-PAH-CT GC column to achieve complete resolution of Chrysene from Triphenylene along with other EU 15+1 PAH compounds. Its unique selectivity helps eliminate false positives while resolving PAH isomers, providing easy, fast, and accurate quantification of PAHs in environmental and food samples.
EU 15+1 PAH Analysis Using Zebron ZB-PAH-CT GC column
| Columns: |
Zebron ZB-PAH-CT |
| Dimension: |
40 meter x 0.18 mm x 0.14 µm |
| Part No.: |
7PD-G044-47 |
| Injection: |
Split 30:1 @ 320 °C, 1 µL |
| Recommended Liner: |
Zebron PLUS Single Taper Z-Liner™ |
| Liner Part No.: |
AG2-4B13-05 (for Shimadzu® 2010 GC) |
| Carrier Gas: |
Helium @ 78 psi (constant pressure) |
| Oven Program: |
45 °C for 0.8 min to 200 °C @ 45 °C/min to 265 °C @ 3 °C/ min for 5 min to 270 °C @ 1 °C/min to 320 °C @ 10 °C/min for 15 min |
| Detector: |
MSD (Shimadzu® GC-MS-QP2010 Ultra) |
| Mode: |
SIM |
| SIM Ions: |
216, 226, 228, 242, 252, 276, 278, 302 m/z |
| Transfer line Temperature: |
300 °C |
| Source Temperature: |
300 °C |
| Sample: |
1. Benzo[c]fluorene 2. Cyclopenta[c,d]pyrene 3. Benz[a]anthracene 4. Triphenylene 5. Chrysene 6. 5-Methylchrysene 7. Benzo[b]fluoranthene 8. Benzo[j]fluoranthene 9. Benzo[k]fluoranthene 10. Benzo[a]pyrene 11. Indeno[1,2,3-c,d]pyrene 12. Dibenz[a,h]anthracene 13. Benzo[g,h,I]perylene 14. Dibenzo[a,l]pyrene 15. Dibenzo[a,e]pyrene 16. Dibenzo[a,I]pyrene 17. Dibenzo[a,h]pyrene |
Shorter Run Time and Better Sensitivity
Comparison of ZB-PAH-CT vs. Popular GC PAH column
Analysis of EU 15+1 and Triphenylene PAHs