Upgrade to Zebron from traditional phases used for PAHs
| Agilent® | Restek® |
|---|---|
| DB®-EUPAH | RXI®-PAH |
Complete Resolution of EU 15+1 and EPA 610 PAHs
Zebron ZB-EU-PAH GC column demonstrates excellent resolution and accurate quantitation of European regulated EU 15+1 and EPA 610 PAHs.
Analysis of EU 15+1 PAHs
|
Sample:
1. Benzo[c]fluorene 2. Benz[a]anthracene 3. Cyclopenta[c,d]pyrene 4. Chrysene 5. 5-Methylchrysene 6. Benzo[b]fluoranthene 7. Benzo[k]fluoranthene 8. Benzo[j]fluoranthene 9. Benzo[a]pyrene 10. Indeno[1,2,3-cd]pyrene 11. Dibenz[a,h]anthracene 12. Benzo[g,h,I]perylene 13. Dibenzo[a,l]pyrene 14. Dibenzo[a,e]pyrene 15. Dibenzo[a,i]pyrene 16. Dibenzo[a,h]pyrene
|
Analysis of EU 15+1 and EPA 610 PAHs
GC-MS conditions for both applications:
|
Column:
Zebron ZB-PAH-EU
|
|
|
Dimension:
30 meter x 0.25 mm x 0.20 µm
|
|
|
Part No.:
7HG-G043-10
|
|
|
Injection:
Split 5:1 @ 330 °C, 1 μL
|
|
|
Recommended Linear:
Zebron PLUS Single Taper Z-Liner™
|
|
|
Liner Part No.:
AG2-4B13-05 (for Shimadzu® 2010 GC)
|
|
|
Carrier Gas:
Helium @ 24 psi (constant pressure)
|
|
|
Oven Program:
45 °C for 0.8 min to 200 °C @ 45 °C/min to 226 °C @ 3 °C/min for 0 min to 320 °C @ 10 °C/min for 20 min
|
|
|
Detector:
MSD, 50-500 m/z |
|
|
Transfer line Temperature:
300 °C |
|
|
Source Temperature:
300 °C |
|
Sample:
1. Naphthalene 2. Acenaphthylene 3. Acenaphthene 4. Fluorene 5. Phenanthrene 6. Anthracene 7. Fluoranthene 8. Pyrene 9. Benzo[c]fluorene 10. Benz[a]anthracene 11. Cyclopenta[c,d]pyrene 12. Chrysene 13. 5-Methylchrysene 14. Benzo[b]fluoranthene 15. Benzo[k]fluoranthene 16. Benzo[j]fluoranthene 17. Benzo[a]pyrene 18. Indeno[1,2,3-cd]pyrene 19. Dibenz[a,h]anthracene 20. Benzo[g,h,I]perylene 21. Dibenzo[a,l]pyrene 22. Dibenzo[a,e]pyrene 23. Dibenzo[a,I]pyrene 24. Dibenzo[a,h]pyrene
|
Up to 70 % Faster GC Analysis
Zebron ZB-PAH-EU will allow you to gain back your precious analysis time! You can easily optimize your column dimensions for greater speed and lab productivity.