Upgrade to Zebron from any 5 %-phenyl or 5 % phenyl-arylene / 95 % dimethylpolysiloxane phase:
| Agilent® | Restek® | Supelco® | |
|---|---|---|---|
| DB®-5ms DB-5ms Ultra Inert DB-5.625 DB-UI 8270D | HP-5ms HP-5ms Ultra Inert VP-5ms CP-5il 8 CB MS | Rxi®-5Sil MS Rxi-5ms | SLB®-5ms |
Enviro-Inert™ Technology: A New Generation In Environmental Testing
Why Is Reduced Activity Important?
Poor inertness as a result of increased column activity can lead to low acid/base sensitivity or analyte misidentification, causing incorrect data and big headaches! ZB-SemiVolatiles is designed with new Enviro-Inert technology to ensure:
- Inert, rugged performance without compromising separation
- Improved resolution of key critical pairs like Benzo[b]fluoranthene and Benzo[k]fluoranthene
- Better peak shapes and response for acids, amines, and PAHs
Enviro-Inert Technology Improves Inertness Without Changing Selectivity
You Spoke, You Tested, You Approved
Your input fueled the research and development of Zebron ZB-SemiVolatiles – the column specifically designed to overcome your EPA Method 8270D obstacles. Several environmental labs have verified real-world performance on ZB-SemiVolatiles:
- TestAmerica Laboratories, Inc. Buffalo
- Phoenix Environmental Laboratories, Inc.
- Other labs like yours!
“Zebron ZB-SemiVolatiles is a very stable and durable semi-volatile column. This has reduced TestAmerica’s downtime and increased our productivity, enabling us to better serve our clients’ needs.”
David Wilkes, GC-MS Semi-Volatile Department TestAmerica Laboratories, Inc. Buffalo
The opinions stated herein are solely those of the speaker and not necessarily those of any company or organization.
We QC Test for the SVOC Compounds You Analyze
We take the guesswork out of meeting method requirements by aggressively testing ZB-SemiVolatiles with two different test mixes. We incorporated troublesome analytes from your samples (Pyridine) and compounds in the 8270D tuning standard (DDT, Pentachlorophenol, and Benzidine) into our QC test, so you can be sure your column is ready to meet suitability requirements for the method.
| Test Probe / Criteria | Pyridine (PYR) Very active amine – exposes any basic activity | |
|---|---|---|
| EPA 8270D Requirement | Our Requirement | |
| Peak Response: | Not Specified | ≥ 0.6* |
| Pentachlorophenol (PCP) PCP peaks disappear and exhibit tailing on columns with acid activity | ||
| EPA 8270D Requirement | Our Requirement | |
| Peak Skew: | ≤ 2.0 | ≤ 2.0 |
| Peak Response: | Not Specified | ≥ 0.3 |
| Benzidine Active amine that tails when basic activity is present | ||
| EPA 8270D Requirement | Our Requirement | |
| Peak Skew: | ≤ 2.0 | ≤ 2.0 |
| DDT Breaks down in an active system to DDE & DDD |